Questions and Answers ​in MRI
  • Home
  • Complete List of Questions
  • ...MR Scanners & Hardware
    • Basic Electromagnetism >
      • What causes magnetism?
      • Radiofrequency waves?
      • Wave properties?
      • What is a tesla?
      • Who was Tesla?
      • What is a gauss?
      • How strong is 3.0T?
      • What is a gradient?
      • Aren't gradients coils?
      • What is susceptibility?
      • Susceptibility causes?
      • How to levitate a frog?
      • What is ferromagnetism?
      • Superparamagnetism?
    • MR Magnets - Part I >
      • Types of magnets?
      • Brands of scanners?
      • Direction of field?
      • Which is the north pole?
      • Low v mid v high field?
      • Advantages to low-field?
      • Disadvantages?
      • What is homogeneity?
      • Why homogeneity?
      • Why shimming?
      • Magnetic field mapping?
      • Passive shimming?
      • Active shimming?
    • MR Magnets - Part II >
      • Superconductivity?
      • Perpetual motion?
      • How to ramp?
      • Superconductive design?
      • Room Temp supercon?
      • Liquid helium use?
      • What is a quench?
      • Is field ever turned off?
      • Emergency stop button?
    • Gradients >
      • Many kinds of coils?
      • Gradient coils?
      • How do z-gradients work?
      • X- and Y- gradients?
      • Open scanner gradients?
      • Eddy current problems?
      • Active shielded gradients?
      • Active shield confusion?
      • What is pre-emphasis?
      • Gradient heating?
      • Gradient specifications?
      • Gradient linearity?
      • Insertable gradients?
    • Radiofrequency Coils >
      • RF Coil function(s)?
      • Radiofrequency waves?
      • RF-transmit coils?
      • LP vs CP (Quadrature)?
      • Multi-transmit RF?
      • Receive-only coils?
      • Array coils?
      • AIR Coils?
    • Site Planning >
      • MR system layout?
      • What are fringe fields?
      • How to reduce fringe?
      • Magnetic shielding?
      • Need for vibration testing?
      • What's that noise?
      • Why RF Shielding?
      • Wires/tubes thru wall?
  • ...The NMR Phenomenon
    • Spin >
      • What is spin?
      • Why I = ½, 1, etc?
      • Proton = nucleus = spin?
      • Predict nuclear spin (I)?
      • Magnetic dipoles?
      • Gyromagnetic ratio (γ)?
      • "Spin" vs "Spin state"?
      • MIsleading diagrams?
      • Quantum "reality"?
      • Zeeman splitting?
      • Fall to lowest state?
      • Net/Bulk magnetization?
      • Does M instantly appear?
    • Precession >
      • Why precession?
      • But nuclei aren't tops!
      • Who was Larmor?
      • Energy for precession?
      • Chemical shift?
      • Does M also precess?
      • Does precession = NMR?
      • Rotating frame?
    • Resonance >
      • Who discovered NMR?
      • MR vs MRI vs NMR?
      • Why is RF input needed?
      • Why Larmor frequency?
      • Off-resonance?
      • CW v Pulsed NMR?
      • How does RF tip M?
      • What is flip angle?
      • Phase coherence?
      • RF-energy absorption?
      • Where is energy stored?
      • Effect of a −90° pulse?
      • RF Energy release?
      • 180° and 360° pulses?
    • NMR Signals >
      • Origin of MR signal?
      • Not radio waves?
      • Continuous wave NMR?
      • Free Induction Decay?
      • Gradient echo?
      • Why not just use the FID?
      • Partial flip angles?
      • How is signal higher?
      • Optimal flip angle?
      • TR and TE?
      • Spin echo?
      • 90°-90° Hahn Echo?
      • Stimulated echoes?
      • STEs for imaging?
      • 4 or more RF-pulses?
    • Relaxation: Physics >
      • Bloch equations?
      • What is T1?
      • What is T2?
      • Relaxation rate vs time?
      • Why is T1 ≥ T2?
      • T2 vs T2*?
      • Causes of Relaxation?
      • Dipole-dipole relaxation?
      • Chemical Exchange?
      • Spin-Spin interactions?
      • Macromolecule effects?
      • Which H's make signal?
      • "Invisible" protons?
      • Magnetization Transfer?
      • T1-rho?
      • Bo effect on T1 & T2?
      • How to predict T1 & T2?
    • Relaxation - Clinical >
      • Actual T1 and T2 values?
      • Meaning of weighting?
      • Spin density [H] effects?
      • Opposite T1 & T2 effects?
      • Long T1 & T2 materials?
      • Short T1 tissues/materials >
        • Adipose tissue?
        • Other fats & oils?
        • Cholesterol?
        • Meconium?
        • Microcalcifications?
        • Melanin?
        • Mucin & colloids?
        • Myelin?
      • Magic angle?
      • MT Imaging/Contrast?
  • ...Making an Image
    • From Signals to Images >
      • Signal squiggles?
      • Real v Imaginary
      • Fourier Transform (FT)?
      • 2D- and 3D-FTs?
      • Who invented MRI?
      • How to locate signals?
    • Frequency Encoding >
      • Frequency encoding?
      • Receiver bandwidth?
      • Narrow bandwidth?
      • Slice-selective excitation?
      • Slice thickness?
      • SS rephasing lobe?
      • Cross-talk?
      • Frequency encode all?
      • Mixing of slices?
      • Two slices at once?
      • Simultaneous Multi-Slice?
    • Phase Encoding >
      • Phase-encoding gradient?
      • Single PE step?
      • What is phase-encoding?
      • PE and FE together?
      • 2DFT reconstruction?
      • PE/FE direction choice?
    • RF & Gradient Details >
      • MR timing diagram?
      • Crushers vs spoilers?
      • CW vs Pulsed RF?
      • RF pulses by function?
      • RF pulse shapes?
      • Adiabatic pulses?
      • Meaning of adiabatic?
      • Use of adiabatic pulses?
      • Adiabatic excitation?
      • Adiabatic pulse design?
      • Variable Rate pulses?
    • Performing an MR Scan >
      • What are the steps?
      • Automatic prescan?
      • Routine shimming?
      • Coil tuning/matching?
      • Center frequency?
      • Transmitter gain?
      • Receiver gain?
      • Dummy cycles?
      • Where's my data?
      • MR Tech qualifications?
    • Image Quality Control >
      • Who regulates MRI?
      • Who accredits?
      • Mandatory accreditation?
      • Routine quality control?
      • MRI Phantoms?
      • Geometric accuracy?
      • Image uniformity?
      • Slice parameters?
      • Image resolution?
      • Signal-to-noise?
      • Ghosting?
  • ...Pulse Sequences
    • Spin Echo >
      • SE vs Multi-SE vs FSE?
      • Image contrast: TR/TE?
      • Does SE correct for T2?
      • Effect of 180° on Mz?
      • Direction of 180° pulse?
    • Inversion Recovery >
      • What is IR?
      • Why use IR?
      • Phase-sensitive IR?
      • Why not PSIR always?
      • TI to null a tissue?
      • Choice of IR parameters?
      • STIR?
      • T1-FLAIR
      • T2-FLAIR?
      • IR-prepped sequences?
      • Double IR?
    • Gradient Echo >
      • GRE vs SE?
      • Multi-echo GRE?
      • GRE sequences?
      • Commercial Acronyms?
      • Spoiling - methods?
      • Spoiled-GRE?
      • Spoiled for T1W only?
      • What is SSFP?
      • FFE/FISP: overview
      • GRASS/FISP: parameters
      • Single- v multi-slice GRE?
      • PSIF vs FISP?
      • True FISP/FIESTA?
      • FIESTA v FIESTA-C?
      • DESS?
      • MERGE/MEDIC?
      • GRASE?
      • MP-RAGE v MR2RAGE?
    • Susceptibility Imaging >
      • What is susceptibility (χ)?
      • What's wrong with GRE?
      • Making an SW image?
      • Phase of blood v Ca++?
      • QSM?
    • Diffusion: Basic >
      • What is diffusion?
      • Iso-/Anisotropic diffusion?
      • "Apparent" diffusion?
      • Making a DW image?
      • What is the b-value?
      • b0 vs b50?
      • Trace vs ADC map?
      • Light/dark reversal?
      • T2 "shine through"?
      • Exponential ADC?
      • T2 "black-out"?
      • ↓ADC: brain?
      • ↓ADC: body?
      • ↓ADC spleen?
    • Diffusion: Advanced >
      • Diffusion Tensor?
      • DTI (tensor imaging)?
      • Tables of DTI Values?
      • Tractography methods?
      • Uses of Tractography?
      • Tractography limitations?
      • Whole body DWI?
      • Small FOV DWI?
      • RO-segmented DWI?
      • IVIM?
      • Diffusion Kurtosis?
    • Fat-Water Imaging >
      • Fat & Water properties?
      • F-W chemical shift?
      • In-phase/out-of-phase?
      • Best method?
      • Dixon method?
      • "Fat-sat" pulses?
      • Water excitation?
      • STIR?
      • SPIR?
      • SPAIR v SPIR?
      • SPIR/SPAIR v STIR?
  • ...K-space & Rapid Imaging
    • K-space: Basic >
      • What is k-space?
      • Parts of k-space?
      • What does "k" stand for?
      • Spatial frequencies?
      • Locations in k-space?
      • Data for k-space?
      • Why signal ↔ k-space?
      • Spin-warp imaging?
      • Big spot in middle?
      • K-space trajectories?
      • Radial sampling?
      • Stack of Stars?
      • GRASP?
    • K-space: Advanced >
      • K-space grid?
      • Negative frequencies?
      • Field-of-view (FOV)
      • Rectangular FOV?
      • Partial Fourier?
      • Phase symmetry?
      • Read symmetry?
      • Why not use both?
      • ZIP?
    • Rapid Imaging (FSE/EPI) >
      • What is FSE/TSE?
      • FSE parameters?
      • Bright Fat?
      • Other FSE differences?
      • Dual-echo FSE?
      • Driven equilibrium?
      • Reduced flip angle FSE?
      • Hyperechoes?
      • SPACE/CUBE/VISTA?
      • Echo-planar imaging?
      • HASTE/SS-FSE?
    • Parallel Imaging >
      • What is PI?
      • How is PI different?
      • PI coils and sequences?
      • Why and when to use?
      • Two types of PI?
      • SENSE/ASSET?
      • GRAPPA/ARC?
      • CAIPIRINHA?
      • Compressed sensing?
      • Noise in PI?
      • Artifacts in PI?
  • ...Contrast Agents
    • Contrast Agents: Physics >
      • Why Gadolinium?
      • Paramagnetic relaxation?
      • What is relaxivity?
      • Why does Gd shorten T1?
      • Does Gd affect T2?
      • Gd & field strength?
      • Best T1-pulse sequence?
      • Triple dose and MT?
      • Dynamic CE imaging?
      • Gadolinium on CT?
    • Contrast Agents: Clinical >
      • So many Gd agents!
      • Important properties?
      • Ionic v non-ionic?
      • Intra-articular/thecal Gd?
      • Gd liver agents (Eovist)?
      • Mn agents (Teslascan)?
      • Blood pool (Ablavar)?
      • Iron-based contrast?
      • Feridex & Liver Agents?
      • Lymph node agents?
      • Ferumoxytol?
      • Ferumoxytol relaxivity?
      • Ferabright™?
      • Bowel contrast agents?
    • Contrast Agents: Safety >
      • Gadolinium safety?
      • Allergic reactions?
      • Renal toxicity?
      • What is NSF?
      • NSF by agent?
      • Informed consent for Gd?
      • Gd protocol?
      • Is Gd safe in infants?
      • Reduced dose in infants?
      • Gd in breast milk?
      • Gd in pregnancy?
      • Gd accumulation?
      • Gd deposition disease?
      • Ferumoxytol Safety?
  • …Cardiovascular and MRA
    • Flow Effects in MRI >
      • Defining flow?
      • Expected velocities?
      • Laminar v turbulent?
      • Predicting MR of flow?
      • Time-of-flight effects?
      • Spin phase effects?
      • Flow void?
      • Why GRE ↑ flow signal?
      • Slow flow v thrombus?
      • Even-echo rephasing?
      • Flow-compensation?
      • Flow misregistration?
    • MR Angiography - I >
      • MRA methods?
      • Time-of-Flight MRA?
      • 2D vs 3D MRA?
      • MRA parameters?
      • Magnetization Transfer?
      • Ramped flip angle?
      • MOTSA?
      • Fat-suppressed MRA?
      • TOF Artifacts
      • Phase-contrast MRA?
      • What is VENC?
      • Measuring flow?
      • 4D Flow Imaging?
      • How accurate?
    • MR Angiography - II >
      • Gated 3D FSE MRA?
      • 3D FSE MRA steps?
      • SSFP MRA?
      • Slab-selective IR MR?
      • MRA with ASL?
      • Dark blood imaging?
      • Vessel wall imaging?
      • Other MRA methods?
      • Contrast-enhanced MRA?
      • Timing the bolus?
      • View ordering in MRA?
      • Bolus chasing?
      • TRICKS or TWIST?
      • MRA artifacts (CE)?
    • Cardiac I: Intro/Anatomy >
      • Cardiac protocols?
      • Patient prep?
      • ECG problems?
      • Magnet changes ECG?
      • Gating: pro- vs retro-?
      • Prospective gating?
      • Retrospective gating?
      • Heart navigators?
      • "Remote" cardiac gating?
      • Dark blood/Double IR?
      • Why not single IR?
      • Triple IR?
      • Polar plots?
      • Coronary artery MRA?
      • Cardiac AI?
      • Cardiac MR artifacts?
    • Cardiac II: Function >
      • Beating heart movies?
      • Cine parameters?
      • Real-time cine?
      • Ventricular function?
      • Cardiac strain?
      • Tagging/SPAMM?
      • Strain Encoding (SENC)?
      • Perfusion: why and how?
      • 1st pass perfusion?
      • Quantifying perfusion?
      • Dark rim artifact
    • Cardiac III: Viability >
      • Gd enhancement?
      • TI to null myocardium?
      • PS (phase-sensitive) IR?
      • Wideband LGE?
      • T1 mapping?
      • ECV Mapping?
      • Iron/T2*-mapping?
      • Edema/T2-mapping?
      • T1-rho mapping?
      • Why/how stress test?
      • Stess drugs/agents?
      • Stress consent form?
  • …Functional Imaging
    • Perfusion I: Intro & DSC >
      • Measuring perfusion?
      • BF, BV, & MTT meaning?
      • DSC v DCE v ASL?
      • How to perform DSC?
      • Bolus Gd effect?
      • T1 effects on DSC?
      • DSC recirculation?
      • DSC curve analysis?
      • DSC signal v [Gd]
      • Arterial input (AIF)?
      • Quantitative DSC?
      • Tmax v TTP v MTT?
    • Perfusion II: DCE >
      • What is DCE?
      • How is DCE performed?
      • How is DCE analyzed?
      • Breast DCE?
      • DCE signal v [Gd]
      • DCE tissue parmeters?
      • K-trans = permeability?
      • Parameters to images?
      • Utility of DCE?
    • Perfusion III: ASL >
      • What is ASL?
      • ASL methods overview?
      • CASL?
      • PASL?
      • pCASL?
      • ASL parameters?
      • ASL artifacts?
      • Gadolinium and ASL?
      • Vascular color maps?
      • Quantifying flow?
    • fMRI/BOLD I: Basic >
      • Who invented fMRI?
      • How does fMRI work?
      • BOLD contrast?
      • Why does BOLD ↑ signal?
      • BOLD=brain activity?
      • BOLD pulse sequences?
      • Task-fMRI Paradigms?
      • Why "on-off" comparison?
      • Task-fMRI: motor?
      • Task-fMRI: visual?
      • Task-fMRI: language?
      • Resting-state fMRI?
      • Resting-state networks?
      • Human "Connectome"?
      • Multi-site studies?
      • fMRI at 7T?
    • fMRI/BOLD II: Advanced >
      • Process/analyze fMRI?
      • Best fMRI software?
      • Data pre-processing?
      • Rs-MRI preprocessing?
      • Registration/normalization?
      • fMRI statistical analysis?
      • General Linear Model?
      • Activation "blobs"?
      • False activation?
      • Multiple comparisons?
      • Mult comparisons control?
      • Analyze RS-fMRI?
      • Correlation analysis?
      • Ind. Component Analysis?
      • ReHo and ALFF?
      • Network/Graphs?
      • Mind reading?
      • fMRI controversies?
  • …MR Artifacts
    • Tissue-related artifacts >
      • Chemical shift artifact?
      • Chemical shift in phase?
      • Reducing chemical shift?
      • Chemical Shift 2nd Kind?
      • In-phase/out-of phase?
      • IR bounce point?
      • Susceptibility artifact?
      • Metal suppression?
      • Dielectric effect?
      • Dielectric Pads?
    • Motion-related artifacts >
      • Why discrete ghosts?
      • Motion artifact direction?
      • Motion reduction?
      • Saturation pulses?
      • Gating methods?
      • Respiratory comp?
      • Camera-based systems?
      • Navigator echoes?
      • Remote resp. gating?
      • PROPELLER/BLADE?
      • SAMER?
      • Motion reduction: AI?
    • Technique-based artifacts >
      • Partial volume effects?
      • Slice overlap?
      • Aliasing?
      • Wrap-around artifact?
      • Eliminate wrap-around?
      • Phase oversampling?
      • Frequency wrap-around?
      • Spiral/radial artifacts?
      • Gibbs artifact?
      • Nyquist (N/2) ghosts?
      • Zipper artifact?
      • Data artifacts?
      • Surface coil flare?
      • MRA Artifacts (TOF)?
      • CE MRA artifacts?
  • …MR Spectroscopy
    • MRS I - Basics >
      • MRI vs MRS?
      • Spectra vs images?
      • Chemical shift (δ)?
      • Measuring δ?
      • Backward δ scale?
      • Predicting δ?
      • Size/shapes of peaks?
      • J-coupling?
      • Localization methods?
      • Single v multi-voxel?
      • PRESS?
      • STEAM?
      • ISIS?
      • CSI?
    • MRS II - Clinical ¹H MRS >
      • How-to: brain MRS?
      • Water suppression?
      • Fat suppression?
      • Normal brain spectra?
      • Choice of TR/TE/etc?
      • Hunter's angle?
      • Lactate inversion?
      • Metabolite mapping?
      • Metabolite quantitation?
      • Breast MRS?
      • Gd effect on MRS?
      • How-to: prostate MRS?
      • Prostate spectra?
      • Muscle ¹H-MRS?
      • Liver ¹H-MRS?
      • MRS artifacts?
    • MRS III - Multi-nuclear >
      • Other nuclei?
      • Why phosphorus?
      • How-to: ³¹P MRS
      • Normal ³¹P spectra?
      • Organ differences?
      • ³¹P measurements?
      • Decoupling?
      • NOE?
      • Carbon MRS?
      • Sodium imaging?
      • Xenon imaging?
  • ...Artificial Intelligence
    • AI Part I: Basics >
      • Artificial Intelligence (AI)?
      • What is a neural network?
      • Machine Learning (ML)?
      • Shallow v Deep ML?
      • Shallow networks?
      • Deep network types?
      • Data prep and fitting?
      • Back-Propagation?
      • DL 'Playground'?
    • AI Part 2: Advanced >
      • What is convolution?
      • Convolutional Network?
      • Softmax?
      • Upsampling?
      • Limitations/Problems of AI?
      • Is the Singularity near?
    • AI Part 3: Image processing >
      • AI in clinical MRI?
      • Super-resolution?
  • ...Tissue Properties Imaging
    • MRI of Hemorrhage >
      • Hematoma overview?
      • Types of Hemoglobin?
      • Hyperacute/Oxy-Hb?
      • Acute/Deoxy-Hb?
      • Subacute/Met-Hb?
      • Deoxy-Hb v Met-Hb?
      • Extracellular met-Hb?
      • Chronic hematomas?
      • Hemichromes?
      • Ferritin/Hemosiderin?
      • Subarachnoid blood?
      • Blood at lower fields?
    • Cartilage mapping?
    • MR Elastography?
    • Synthetic MRI?
    • Amide Proton Transfer?
    • MR thermography?
    • MR Fingerprinting?
    • Electric Properties Imaging?
  • ...Safety and Screening
    • Overview >
      • Status of the 5G line?
      • ACR Safety Zones?
      • MR safety screening?
      • Incomplete screening?
      • Passive/active implants?
      • Conditional implants?
      • Common safety issues?
      • Projectiles?
      • Metal detectors?
      • Pregnant patients?
      • Postop/ER/ICU patients?
      • Temperature monitoring?
      • Orbital foreign bodies?
      • Bullets and shrapnel?
    • Static Fields >
      • "Dangerous" metals?
      • "Safe" metals?
      • Magnetizing metal?
      • Object shape?
      • Forces on metal?
      • Most dangerous place?
      • Force/torque testing?
      • Static field bioeffects?
      • Dizziness/Vertigo?
      • Flickering lights?
      • Metallic taste?
    • RF Fields >
      • RF safety overview?
      • RF biological effects?
      • What is SAR?
      • SAR limits?
      • How MRIs measure SAR?
      • Operating modes?
      • How to reduce SAR?
      • RF burns?
      • Estimate implant heating?
      • SED vs SAR?
      • B1+rms vs SAR?
      • Personnel exposure?
      • Cell phones?
    • Gradient Fields >
      • Gradient safety overview
      • Acoustic noise?
      • Nerve stimulation?
      • Gradient vs RF heating?
    • Safety: Neurological >
      • Aneurysm coils/clips?
      • Shunts/drains?
      • Pressure monitors/bolts?
      • Deep brain stimulators?
      • Spinal cord stimulators?
      • Vagal nerve stimulators?
      • Cranial electrodes?
      • Carotid clamps?
      • Peripheral stimulators?
      • Epidural catheters?
    • Safety: Head & Neck >
      • Additional orbit safety?
      • Cochlear Implants?
      • Bone conduction implants?
      • Other ear implants?
      • Dental/facial implants?
      • ET tubes & airways?
    • Safety: Chest & Vascular >
      • Breast tissue expanders?
      • Breast biopsy markers?
      • Airway stents/valves/coils?
      • Respiratory stimulators?
      • Ports/vascular access?
      • Swan-Ganz catheters?
      • IVC filters?
      • Infusion pumps?
      • Insulin pumps & CGMs?
      • Vascular stents/grafts?
      • Sternal wires/implants?
    • Safety: Cardiac >
      • Pacemaker dangers?
      • Pacemaker terminology?
      • New/'Safe" Pacemakers?
      • Old/Legacy Pacemakers?
      • Violating the conditions?
      • Prep/Consent for CIEDs
      • Epicardial pacers/leads?
      • Cardiac monitors?
      • Heart valves?
      • Miscellaneous CV devices?
    • Safety: Abdominal >
      • PIllCam and capsules?
      • Gastric pacemakers?
      • Other GI devices?
      • Contraceptive devices?
      • Foley catheters?
      • Incontinence devices?
      • Penile Implants?
      • Sacral nerve stimulators?
      • GU stents and other?
    • Safety: Orthopedic >
      • Orthopedic hardware?
      • External fixators?
      • Traction and halos?
      • Bone stimulators?
      • Magnetic rods?
  • Copyright/Legal
    • Copyright Issues
    • Legal Disclaimers
  • Forums/Blogs/Links
  • Self-test Quizzes
    • Magnets & Scanners Quiz
    • Safety & Screening Quiz
    • NMR Phenomenon Quiz
    • Pulse Sequences Quiz
    • Making an Image Quiz
    • K-space & Rapid Quiz
    • Contrast & Blood Quiz
    • Cardiovascular & MRA Quiz

T1-FLAIR

What is T1-FLAIR?  
T1-FLAIR
T1-FLAIR stands for T1-weighted-Fluid-Attenuated Inversion Recovery. This nomenclature began to arise in the late 1990s to denote an inversion recovery sequence with dark CSF and other T1-like properties made possible by a medium TI coupled with fast spin-echo signal acquisition.  In reality, this was pretty much the standard IR technique that had been used for years, just tuned up and rebranded for the modern era.
For an IR pulse sequence using fast spin-echo acquisition, signal intensity is nulled when TI is chosen so that the term
T1-FLAIR equation
Because CSF has a long intrinsic T1 value (3000-4000 ms), its TInull is strongly dependent on the TR selected. For T1-FLAIR both the TR and TE values are kept relatively short to minimize T2-weighting. Considering these restrictions, the TInull for a T1-FLAIR sequence is typically in the range of 800-1000 ms. The graph below shows the approximate relationship between TI and TR for fluid/CSF suppression.
T1-FLAIR
Approximate TI required to null CSF signal depends on TR. (Here we assume T1 of CSF = 4000 and fast spin echo readout used with TElast = 60)
T1-FLAIR
T1-FLAIR brain image at 3T with TR=2100, TE=9, TI=880.

Advanced Discussion (show/hide)»

No supplementary material yet. Check back soon!

References
     Bydder GM, Young IR. MR imaging: clinical use of the inversion recovery sequence. J Comput Assist Tomogr 1985; 9:659-675.     
     Melhem ER, Israel DA, Eustace S, Jara H. MR of the spine with a fast T1-weighted fluid-attenuated inversion recovery sequence. AJNR Am J Neuroradiol 1997; 18:447-454.

Related Questions
     How do you pick a TI value to null signal from a given tissue? 
     What is T2-FLAIR?

←  Previous Question
Next Question  →